Ultra-high data throughput: 14 Tb/Run
Ultra-high production efficiency: 9 Tb/day, over 23,000 WGS (30x)/ year
Innovative sequencing chemistry and Al base-calling algorithm
Q40 level data output can be achieved
Fully automated independent eight-channel sample loading
Supports break-point resumption for worry-free project operation
Customizable sequencing order and data batch delivery
Automatic cleaning after unloading and automatic reagent discharge
| SURFSeq Q V1.0 Kit | ||||||||
| Flow Cell | Lane | Throughput(Reads/FC) | Type | Read Length |
Data output1 (FC×2) |
Q301 |
Q401 |
TAT2 (1 FC) |
| FCM | 4 | 11.7B | 50 cycles | SE50 | 0.6 Tb×2 | Q30≥90% | Q40≥90% | 10 hr |
| 100 cycles | SE100/PE50 | 1.2 Tb×2 | 13 hr | |||||
| 200 cycles | PE100 | 2.3 Tb×2 | 19 hr | |||||
| 300 cycles | PE150 | 3.5 Tb×2 | 24 hr | |||||
| FCH | 8 | 23.3B | 50 cycles | SE50 | 1.2 Tb×2 | 17 hr | ||
| 100cycles | SE100/PE50 | 2.3 Tb×2 | 22 hr | |||||
| 200cycles | PE100 | 4.7 Tb×2 | 29 hr | |||||
| 300cycles | PE150 | 7 Tb×2 | 36 hr | |||||
1. This parameter is obtained based on the average of multiple test resuts of GeneMind P3 standard library. The data output and the proportion of high-quality data are affected by factors such as sample type, sample quality and effective flow cell utilization. The actual performance may vary.
2. The TAT (tumaround time)for a standard run (paired-end 8+8 indexing)encompasses the entire process from library loading to the generationtion of base-called sequence files.
| Applications | Read Length | Data/sample | Number of samples/FC | |
| FCM-11700M | FCH-23400M | |||
| Oncology Panel (580genes)-Plasma | PE150 | 50 Gb | 48 | 96 |
| WES (100X) | PE150 | 12 Gb | 256 | 512 |
| WGS (30X) | PE150 | 110 Gb | 26 | 52 |
| Single cell 3' RNA-seq | PE150 | 120 Gb | 24 |
48 |
The above sample numbers for different applications are only for reference. User need to adjust the sample numbers according to the actual experiment.
| Specification | |
| Dimensions W × D × H | 975 mm × 921 mm×1570 mm |
| Weight | < 500 kg |
| Power requirements | < 6000 W |
| 200-240 V~,50/60 Hz | |
| Operating environment | Temperature: 19 ℃ – 25 ℃ |
| non-condensing:20‒80% relative humidity | |
| Instrument control computer | Altitude:≤2000 m |
| CPU:Intel Xeon Gold 6326 ×2 | |
| Memory: 2 TB | |
| Solid-state drive: 3.7 TB | |
| Hard disk drive: 76.3 TB | |
| Operating system: Windows 10 ×64 | |
| Catalog Number | Product Name | Specification |
| SQ00066 | SURFSeq Q Sequencing System Set | RUO |
| S000379 | SURFSeq Q Sequencing Kit V1.0 (FCM 50cycles) | FCM 50cycles |
| S000380 | SURFSeq Q Sequencing Kit V1.0 (FCM 100cycles) | FCM 100cycles |
| S000381 | SURFSeq Q Sequencing Kit V1.0 (FCM 200cycles) | FCM 200cycles |
| S000382 | SURFSeq Q Sequencing Kit V1.0 (FCM 300cycles) | FCM 300cycles |
| S000383 | SURFSeq Q Sequencing Kit V1.0 (FCH 50cycles) | FCH 50cycles |
| S000384 | SURFSeq Q Sequencing Kit V1.0 (FCH 100cycles) | FCH 100cycles |
| S000385 | SURFSeq Q Sequencing Kit V1.0 (FCH 200cycles) | FCH 200cycles |
| S000386 | SURFSeq Q Sequencing Kit V1.0 (FCH 300cycles) |
FCH 300cycles |
For Research Use Only. Not intended for use in diagnostic or clinical procedures. Not available in all countries.